Miami J Select Collar, Catalog: MJS-101; Miami J Select Collar Set, Catalog: MJSR-101

Z-0006-2025

August 20, 2024

Class II

Product summary

Firm
Ossur H / F
Event
Event 95278
Status
Ongoing
Classification
Class II
Quantity
21,020
Official record key
device-enforcement:Z-0006-2025

Official wording

Reason: Pressure sensitive adhesive hooks attached to the sides of the anterior cervical orthosis collar panel can occasionally exhibit poor adhesion, which may lead to them gradually peeling off with repeated doffing, possibly leading to reduced immobilization of the cervical spine.

Code information: REF/UDI-DI/GTIN/Lot Range/Sold in the Period, Date Range: MJS-101/05690967817008/MX220516 to MX230104/2022-05-19 to 2023-06-05, MJSR-101/05690967817145/MX220516 to MX230104/2022-05-19 to 2023-06-05

Distribution pattern: US: NH, MA, NY, NJ, PA, GA, MD, NC, IN, FL, AR, TN, KY, OH, MI, IL, TX, MO, CO, UT, AZ, WA, OR, CA, NM, VA, IA, MN, NE, SC, MT, LA, ID, SD, AL, CT, OK, AK, PR

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0