Device product
Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM JAPAN; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which includes intravascular ultrasound (IVUS). When used with a peripheral vascular imaging catheter, the system utilizes the IVUS modality for ultrasound examination of intravascular pathology.
Z-0108-2025
Product summary
- Event
- Event 95331
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 9 units
- Official record key
device-enforcement:Z-0108-2025
Official wording
Reason: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
Code information: Material Number H7492493220C0, GTIN/UDI/DI 00191506033231, Batch/Serial Numbers: 103045130, 103427062, 103427064, 103460623, 103545307, 103545308, 103679278, 103679281, 103680004
Distribution pattern: US Nationwide distribution in Puerto Rico.
Derived failure modes
-
Device software or design
software anomaly