BD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10, 5.20 and 5.30- IVD standalone laboratory software solution providing data management and workflow management functionality across locations where clinical diagnostic activities take place. Catalog Number: 444150
Z-0026-2025
Product summary
- Event
- Event 95394
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 162 units
- Official record key
device-enforcement:Z-0026-2025
Official wording
Reason: Synapsys software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This occurs when the identified organism is included in the same interpretive group as a species for which there is a specific breakpoint. May result in either a false susceptible or false resistant result. A false susceptible result can cause inadequate or incomplete antibiotic treatment due to the use of an ineffective antibiotic, and a false resistant result may limit access to a preferred antibiotic regimen with possible elongation of therapy
Code information: UDI-DI: 00382904441500 Software versions: 4.20, 4.30, 5.10, 5.20 and 5.30
Distribution pattern: Nationwide Foreign: Australia, New Zealand, Singapore, Hong Kong, Taiwan, Japan, Austria, Belgium, Denmark, France, Germany, Italy, Luxembourg, Netherlands, Norway, Spain, Sweden, Switzerland, United Kingdom, Russia, Saudi Arabia, South Africa, Canada, Mexico, China
Derived failure modes
-
Unknown
Synapsys software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This occurs when the identified organism is included in the same interpretive group as a species for which there is a specific breakpoint. May result in either a false susceptible or false resistant result. A false susceptible result can cause inadequate or incomplete antibiotic treatment due to the use of an ineffective antibiotic, and a false resistant result may limit access to a preferred antibiotic regimen with possible elongation of therapy