Skip to content
Recall Observatory FDA recall evidence

Device product

BARD MAGNUM Disposable Core Tissue Biopsy Needle REF: MN1410 MN 1420 MN1610 MN1616 MN1820 MN2010 The Magnum Biopsy System (instrument and needles) is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, breast, spleen, lymph nodes and various soft tissue tumors.

Z-3206-2024

August 02, 2024

Class II

Product summary

Firm
Bard Peripheral Vascular Inc
Event
Event 95174
Status
Ongoing
Classification
Class II
Quantity
60 (U.S.) and 5,160 (O.U.S.)
Official record key
device-enforcement:Z-3206-2024

Official wording

Reason: Due to possible packaging damage (indentations/cuts) cause by manufacturing process. As a result of the packaging damage, may cause the sterile barrier to be compromised.

Code information: Product Name: Magnum 14g x 10cm Catalog Number: MN1410 Lot Numbers / UDI code: REJP3295 / (01)00801741084249(17)270131(10)REJP3295 REJP2228 / (01)00801741084249(17)270131(10)REJP2228 REJN2583 / (01)00801741084249(17)261231(10)REJN2583 REJQ2851 / (01)00801741084249(17)270228(10)REJQ2851 REJQ0831 / (01)00801741084249(17)270228(10)REJQ0831 Product Name: Magnum 14g x 20cm Catalog Number: MN1420 Lot Numbers / UDI code: REHX2867 / (01)00801741084270(17)260930(10)REHX2867 Product Name: Magnum 16g x 10cm Catalog Number: MN1610 Lot Numbers / UDI code: REJN0587 / (01)00801741084287(17)261231(10)REJN0587 Product Name: Magnum 16g x 16cm Catalog Number: MN1616 Lot Numbers / UDI code: REJP3289 / (01)00801741084300(17)270131(10)REJP3289 Product Name: Magnum 18g x 20cm Catalog Number: MN1820 Lot Numbers / UDI code: REJS1418 / (01)00801741084355(17)270430(10)REJS1418 Product Name: Magnum 20g x 10cm Catalog Number: MN2010 Lot Numbers / UDI code: REJQ1160 / (01)00801741084386(17)270228(10)REJQ1160

Distribution pattern: Worldwide - U.S. Nationwide distribution in the Puerto Rico and the countries of Canada, Pakistan, Peru, India, Malaysia, China, Brazil, Mexico, Japan, Chile, Australia.

Derived failure modes

  • Sterility assurance

    reason.sterility_assurance · v1.0.0

    sterile barrier to be compromised