Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation, used to provide airway clearance therapy

Z-0208-2025

September 16, 2024

Class I

Product summary

Firm
Percussionaire Corporation
Event
Event 95436
Status
Ongoing
Classification
Class I
Quantity
43,830 devices
Official record key
device-enforcement:Z-0208-2025

Official wording

Reason: As a result of product complaints, breathing circuits are experiencing discolored fluid which are formed when therapy was being given with specific medication through the nebulizer cup. Discoloration is caused by a reaction between medicated mist and an inadvertent/unapproved addition of a nickel coating on the spring component.

Code information: Model Numbers: P5-10 P5-HC P5-UC-10 UDI-DI codes: Lot Numbers: 240321 240329 240412 240430 240509 240516 240614 240625 240627

Distribution pattern: U.S.: PA, AR, MI, CA, TX, OH, NY, FL, IL, NE, OR, GA, WI, MO, OK, AL, WV, MT, CO, LA, VA, CT, SC, NH, NV, NC, NJ, MI, UT, ID, ME, MN, SD, ID, ND, KS, KY, TN, MD, VT, MA, RI, O.U.S.: Canada, Switzerland (all products shipped OUS remained in Sentec AG's (distributor, parent company of Percussionaire) control, were quarantined, and did not leave the warehouse).

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    As a result of product complaints, breathing circuits are experiencing discolored fluid which are formed when therapy was being given with specific medication through the nebulizer cup. Discoloration is caused by a reaction between medicated mist and an inadvertent/unapproved addition of a nickel coating on the spring component.