Skip to content
Recall Observatory FDA recall evidence

Device product

medex TranStar KIDS KIT 7INCH CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9547

Z-0200-2025

September 24, 2024

Class II

Product summary

Firm
Smiths Medical ASD, Inc.
Event
Event 95438
Status
Ongoing
Classification
Class II
Quantity
3280 units
Official record key
device-enforcement:Z-0200-2025

Official wording

Reason: Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

Code information: UDI/DI 50351688503013 (case), 10351688503015 (pouch), Lot Numbers: 4223699, 4248326, 4297927, 4323137, 4340498, 4346068, 4350022

Distribution pattern: Worldwide distribution.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    manufacturing defect