Skip to content
Recall Observatory FDA recall evidence

Device product

Soltive Pro SuperPulsed Laser Systems-Intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue Model/Catalog Number: TFL-SLS

Z-0135-2025

August 13, 2024

Class II

Product summary

Firm
Olympus Corporation of the Americas
Event
Event 95380
Status
Ongoing
Classification
Class II
Quantity
159 units
Official record key
device-enforcement:Z-0135-2025

Official wording

Reason: The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.

Code information: UDI 00821925044135 All Serial Numbers

Distribution pattern: Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.