Skip to content
Recall Observatory FDA recall evidence

Device product

Optima MR360, with affected software versions: SV20.3, SV23.3, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System

Z-2478-2024

June 14, 2024

Class II

Product summary

Firm
GE Medical Systems, LLC
Event
Event 94874
Status
Ongoing
Classification
Class II
Quantity
469 units
Official record key
device-enforcement:Z-2478-2024

Official wording

Reason: GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.

Code information: System ID: 1009000, 201871MR, DZ1396MR01, DZ1820MR01, DZ1339MR02, DZ1339MR01, DZ1061MR01, DZ1406MR01, DZ1285MR01, DZ1191MR01, 0910272060, 0910274038, 0910274036, BH1038MR01, MRR9112, MRR10326, MRR8777, MRR10248, MRR9308, MRR10649, MRR10657, MRR10586, MRR10646, MRR10703, MRR10694, MRR10648, MRR10676, MRR8247, MRR10417, MRR10842, MRR10620, MRR10145, MRR9922, MRR9979, MRR10393, MRR9829, MRR8339, MRR11105, MRR10416, MRR10514, MRR10812, MRR9091, MRR10742, MRR8318, MRR10418, MRR9927, MRR8632, MRR8679, MRR8213, MRR9379, MRR10692, MRR10259, MRR10205, MRR10330, MRR11034, MRR9826A, MRR10690, MRR9033, MRR10478, MRR10529, MRR8249, MRR9080, MRR8969, MRR10201, MRR8714, MRR10793, MRR10013A, MRR8427, MRR11089, MRR10594, MRR9435, MRR10177, MRR10629, MRR10366, MRR10640, MRR10596, MRR9988, MRR10158, MRR10486, MRR9587, MRR10341, MRR10388, MRR10147, MRR10206, MRR9797, MRR9636A, MRR9450, MRR9947, MRR9271, MRR10227, MRR10062, MRR9000, MR1551SX1, MRR10336, MRR10115, MRR9877, MRR10726, MR1578SX4, MR1462SX1, MRR9318, MRR10645, MRR10218, MRR10076, MRR8392, MRR8861, MRR8835, MRR8194, MRR10584, MR1532SX1, MRR8424, MRR10442, MRR10885, MRR8415, MRR10603, MRR10606, MRR10619, MR1179SX1, MRR10656, MRR9500, MRR10680, MRR10143, MRR10082, MRR10065, MRR9544, MRR8831, MRR11307, MRR10642, MRR9043, 082427040158, 082427040159, 082427040131, 082427040134, 082427040136, 082427040138, 082427220034, 082427040156, 082427120128, 082427200082, 082427010067, 082427240051, 082427270031, 082427160112, 082427090225, 082427040130, 082427120140, 082427120139, 082427120123, 082427060036, 082427010068, 082427070165, 082427050035, 082427100179, 082427210089, 082427140118, 082427240036, 082427120111, 082427060054, 082427160119, 082427160093, 082427100209, 082427210112, 082427230086, 082427120105, 082427250035, 082427100205, 082427030083, 082427100197, 082427160102, 082427020051, 082427120108, 082427060045, 082427070085, 082427090182, 082427060041, 082427120115, 082427190055, 082427080035, 082427270038, 082427130070, 082427090222, 082427090203, 082427160116, 082427160115, 082427160111, 082427160092, 082427160121, 082427240029, 082427240026, 082427070121, 082427070088, 082427070106, 082427070093, 082427210122, 082427070105, 082427070124, 082427070118, 082427090239, 082427090210, 082427090219, 082427090236, 082427090209, 082427090207, 082427210117, 082427210140, 082427210100, 082427210104, 082427190061, 082427120094, 082427030087, 082427030085, 082427060047, 082427060053, 082427300016, 082427060044, 082427060050, 082427310106, 082427090213, 082427190094, 082427100190, 082427130068, 082427100177, 082427100193, 082427100212, 082427100222, 082427100178, 082427100174, 082427100206, 082427100196, 082427100195, 082427100180, 082427100188, 082427080039, 082427310088, 082427310043, 082427270047, 082427270041, 082427140095, 082427250043, 082427060043, 082427060039, 082427300017, 082427100152, 082427100203, 082427100186, 082427100171, 082427080045, 082427180024, 082427270036, 082427120096, 082427120119, 082427120109, 082427200057, 082427070084, 082427070082, 082427070099, 082427070080, 082427070086, 082427090217, 082427140092, 082427190074, 082427030075, 082427030072, 082427030091, 082427310052, MRR8406, EG1925MR01, EG3053MR01, EG1840MR02, EE4024MR01, 782057MR01, PC2287MR01, PC0395MR02, 805466MR03, 363893MR01, PC4529MR01, 782692MR01, HC7831MR01, 734128MR01, 164236MR02, H4678MR04, MRR8465, MRR8457, 083027105723017, 083027820014211, 083027288790310, 083027201106511, 083027842022620, 083027801024814, 083027808008011, 083027849290410, 083027000013013, 083027600004313, 0835270028, 0835270027, IQ1309MR01, A5714101, A106327806, PC0098, PC0089, PC0107, PC0075, PC0078, PC0048, PC0047, PC0049, PC0071, PC0050, PC0090, PC0115, PC0117, PC0085, PC0096, PC0057, PC0083, PC0087, PC0067, PC0082, PC0046, PC0052, PC0105, PC0110, PC0091, PC0068, PC0133, PC0130, PC0088, PC0122, PC0076, PC0101, PC0084, PC0074, PC0053, PC0086, PC0064, PC0042, PC0032, PC0111, PC0054, PC0103, PC0094, PC5005, PC0061, PC0065, PC0069, PC6102, PC0079, PC0041, PC0120, PC0077, PC0106, PC0066, PC0063, PC0039, PC0036, PC0097, PC0044, PC0051, PC0062, PC0109, PC0123, PC0102, PC0043, PC0035, PC0095, JO1147MR01, 0850270360, 0850270375, 0850270405, 0850270397, 0850270366, 0850270387, 0850270382, 0850270373, 0850270368, LV4012MR02, LV4031MR01, LB1359MR01, LB1088MR02, LB1358MR01, LB1364MR01, LT4026MR01, LT4043MR01, 2836414HDE, MRR8505, MA1134MR01, MA1487MR01, MA1471MR01, NG1126MR01, NO1200MR01, 0843270022, PL1620MR01, PL2059MR01, PL1798MR01, P00571MR01, PL2433MR01, PL1321MR01, PL2435MR01, PL2854MR01, P00607MR01, PL1951MR01, PPR44921, PPS07404, RO1105MR01, RO1223MR01, RF1709MR01, RF2036MR01, RU7690MR01, RU4356MR01, RU1130MR01, RU8412MR01, RU1346MR01, RU1136MR01, RU1875MR01, 600100MR03, RU1146MR01, RU3883MR01, RU1820MR02, SA2305MR01, SA2352MR01, SA2191MR04, SA2253MR01, SA1056MR03, ZA2198MR01, ZA2383MR01, MPX82245, UP925712, NPV86500, EPP04112, DPN95902, DPL32939, MPD24000, GPQ13902, NPV82605, UPW07802, MPX09800, FPG08503, MP720818, MPX75606, DPN79400, MPD19500, DPN62603, EPP11100, SY2060MR01, SY2058MR01, 0853270123, 0853270131, 0853270134, 0853270139, 0853270126, 0856270052, 0856270049, TN1033MR02, TN1241MR01, TN1008MR01, TN1530MR01, 16048MRS01, 16049MRS02, 34337MRS01, 42021MRS01, 68008MRS01, 34346MRS01, AE1373MR01, UZ1337MR01, 3882941838MRCTR, 3882928195MRMET, 0859270015, 0859270014, 0859270009, 0859270008

Distribution pattern: Worldwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.