Ingenia 3.0T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782103

Z-3195-2024

September 03, 2024

Class II

Product summary

Firm
Philips North America Llc
Event
Event 95289
Status
Ongoing
Classification
Class II
Quantity
33 units
Official record key
device-enforcement:Z-3195-2024

Official wording

Reason: Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

Code information: UDI-DI: (01)00884838098299(21) Serial Numbers: 85625 85626 85628 85630 85633 85637 85638 85641 85642 85646 85650 85652 85653 85654 85656 85657 85658 85659 85660 85662 85664 85666 85667 85668 85672 85674 85682 85683 85686 85692 85696 85701 85713

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.