The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.

Z-2337-2024

May 31, 2024

Class II

Product summary

Firm
Maquet Cardiovascular, LLC
Event
Event 94802
Status
Ongoing
Classification
Class II
Quantity
600 units
Official record key
device-enforcement:Z-2337-2024

Official wording

Reason: There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue.

Code information: UDI-DI: 00607567700345 Lot Numbers; 3000371260 3000378288 3000362934 3000371798 3000378554 3000372477 3000378950 3000376542 3000380314 3000377320 3000380847

Distribution pattern: NationwideForeign: AT DE DK ES GB HK IT NL ZA

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue.