Device product
Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
Z-2441-2018
Product summary
- Firm
- Zeiss, Carl Inc
- Event
- Event 80418
- Status
- Terminated
- Classification
- Class II
- Quantity
- 8
- Official record key
device-enforcement:Z-2441-2018
Official wording
Reason: Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.
Code information: 432332-9020-000
Distribution pattern: US Nationwide Distribution
Derived failure modes
-
Unknown
Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.