Skip to content
Recall Observatory FDA recall evidence

Device product

Olympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units to cut tissue using high-frequency current within the upper digestive tract Model: KD-645L

Z-1548-2024

March 01, 2024

Class II

Product summary

Firm
Olympus Corporation of the Americas
Event
Event 94173
Status
Ongoing
Classification
Class II
Quantity
2525 units
Official record key
device-enforcement:Z-1548-2024

Official wording

Reason: Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions for Use, which details critical specifications regarding electrosurgical unit compatibility and output due to reports of serious injury

Code information: UDI-DI: - 04953170407857 All lots numbers

Distribution pattern: Nationwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions for Use, which details critical specifications regarding electrosurgical unit compatibility and output due to reports of serious injury