Device product
Neurostimulation Kit (Ported, Dual 8, 40 cm), REF: 71005-US Peripheral Nerve Stimulation Kit (Ported, Dual 8, 40 cm), REF: 71019-US Components of the Nalu Spinal Cord Stimulation System.
Z-2257-2024
Product summary
- Event
- Event 94682
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 13
- Official record key
device-enforcement:Z-2257-2024
Official wording
Reason: Neurostimulation Systems have a manufacturing non-conformance that may reduce the efficiency of the radiofrequency signals between the Implantable Pulse Generator (IPG) and the Therapy Disc, causing the IPG to be non-functional, which causes the device to be non-functional so it will no longer inhibit the transmission of pain signals to the brain.
Code information: UDI-DI: 71005-US: 00812537036137, 00812537033600, 00812537033518, 00812537030975, 00812537030524. 71019-US: 00812537036441, 00812537035604, 00812537035598 Serial Numbers: H01377, H01382, H01432, H01419, H01387, H01445, H01415, H01441, H01373, H01428, H01378, H01438, H01431
Distribution pattern: US Nationwide distribution in the states of MO, IL, CA, TX, TN, OK, AZ, AR.
Derived failure modes
-
Unknown
Neurostimulation Systems have a manufacturing non-conformance that may reduce the efficiency of the radiofrequency signals between the Implantable Pulse Generator (IPG) and the Therapy Disc, causing the IPG to be non-functional, which causes the device to be non-functional so it will no longer inhibit the transmission of pain signals to the brain.