Skip to content
Recall Observatory FDA recall evidence

Device product

Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the Sight OLO CBC analyzer.

Z-2173-2024

May 13, 2024

Class II

Product summary

Firm
SIGHT DIAGNOSTICS LTD
Event
Event 94678
Status
Ongoing
Classification
Class II
Quantity
7,450 units (2,800 US, 4,650 OUS)
Official record key
device-enforcement:Z-2173-2024

Official wording

Reason: One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results in the test kit lot performing outside of the device specification.

Code information: REF: TK1, UDI-DI: B787TK10, Lot Code: 10114

Distribution pattern: Worldwide - US Nationwide distribution in the states of AL, FL, MS, TX, WI and the countries of Canada, Israel, Slovenia, Trinidad & Tobago.

Derived failure modes

  • Microbial contamination

    reason.microbial_contamination · v1.0.0

    bacterial contamination