Skip to content
Recall Observatory FDA recall evidence

Device product

MEDLINE LUER LOCK DISPOSABLE SYRINGE WITHOUT NEEDLE, REF SYR101010

Z-1937-2024

April 22, 2024

Class II

Product summary

Firm
MEDLINE INDUSTRIES, LP - Northfield
Event
Event 94537
Status
Ongoing
Classification
Class II
Quantity
25,214,334 syringes
Official record key
device-enforcement:Z-1937-2024

Official wording

Reason: It has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range of their claimed graduated capacity and may pose a risk to patient health.

Code information: UDI/DI 10884389596319, All lots in the format: 897YYMMXXXX

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Canada, Panama, UAE, USVI.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    It has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range of their claimed graduated capacity and may pose a risk to patient health.