The VASOVIEW HemoPro Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.

Z-2176-2024

May 17, 2024

Class II

Product summary

Firm
Maquet Cardiovascular, LLC
Event
Event 94700
Status
Ongoing
Classification
Class II
Quantity
37,186 units
Official record key
device-enforcement:Z-2176-2024

Official wording

Reason: Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.

Code information: Model No. VH-3500, VH-3000-W; UDI: 00607567701250, 00607567700345; Lot No. 3000314373, 3000314821, 3000316023, 3000316024, 3000316027, 3000317501, 3000317575, 3000317576, 3000318302, 3000318901, 3000320140, 3000320141, 3000320773, 3000321483, 3000321723, 3000323214, 3000324397, 3000324409, 3000325013, 3000325249, 3000325689, 3000325776, 3000328206, 3000328561, 3000328562, 3000329723, 3000330445, 3000331131, 3000331388, 3000332428, 3000332510, 3000332993, 3000333524, 3000333829, 3000333956, 3000333967, 3000334315, 3000335670, 3000336708, 3000337272, 3000339701, 3000340534, 3000341133, 3000341675, 3000342250, 3000342251, 3000343264, 3000343266, 3000344887, 3000351187, 3000352770, 3000354226, 3000354228, 3000355224, 3000355369, 3000355798, 3000356048, 3000357837, 3000360574, 3000360576, 3000360775, 3000361036, 3000361423, 3000361707, 3000362064, 3000362258, 3000362671, 3000362933, 3000363162, 3000363947, 3000364082, 3000364309, 3000364520, 3000365379, 3000365391, 3000366446, 3000366831, 3000367160, 3000367439, 3000367542, 3000367852, 3000368165, 3000368658, 3000369005, 3000369122, 3000369645, 3000370069, 3000370290, 3000372257, 3000375534, 3000375776, 3000376132, 3000376388, 3000376542, 3000376883, 3000376883, 3000377320, 3000377699, 3000378079, 3000378654, 3000379282, 3000379691, 3000380264, 3000380475, 3000380671, 3000381287, 3000381620, 3000382054, 3000383367, 3000383779, 3000384088, 3000388588.

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Australia Brazil, Canada, China, EEA, India, Korea, New Zealand, Southern Africa, Taiwan, Thailand, Turkey, & United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.