Device product
RayStation 11B
Z-1299-2022
Product summary
- Event
- Event 90505
- Status
- Terminated
- Classification
- Class II
- Quantity
- 26
- Official record key
device-enforcement:Z-1299-2022
Official wording
Reason: This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.
Code information: UDI-DI 0735000201042620211208 Software version numbers 12.0.0.932, 12.1.0.1221, 12.0.3.68
Distribution pattern: US Nationwide distribution in the states of NJ and TX.
Derived failure modes
-
Unknown
This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.