Skip to content
Recall Observatory FDA recall evidence

Device product

Discrete photometric chemistry analyzer for clinical use.

Z-2545-2024

June 14, 2024

Class II

Product summary

Firm
Beckman Coulter, Inc.
Event
Event 94902
Status
Ongoing
Classification
Class II
Quantity
59 units
Official record key
device-enforcement:Z-2545-2024

Official wording

Reason: Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.

Code information: Catalog Number: 81600N, C11907; UDI: 150995901; Serial Numbers: 574415, 574440, 574484,574416, 574441, 574485, 574417, 574442, 574486, 574418, 574443, 574488,574419, 574444, 574489,574420, 574445, 574490, 574421, 574446, 574491, 574422, 574447 574494,574423, 574448, 574497,574424, 574454, 574502, 574425, 574455, 574503, 574426, 574456 574504,574427, 574457, 574505,574428, 574459, 574506, 574429, 574469, 574507, 574430, 574470 574509,574432, 574471, 574513, 574433, 574472, 574514, 574434, 574477, 574515, 574435, 574479

Distribution pattern: Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY, CA, MI, AL, IL the countries of Austria, Belgium, Georgia, India, Indonesia, Italy, Korea, Malaysia, Myanmar, Nigeria, Switzerland, Vietnam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.