Skip to content
Recall Observatory FDA recall evidence

Device product

IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Z-1945-2024

May 10, 2024

Class II

Product summary

Firm
Ion Beam Applications S.A.
Event
Event 94576
Status
Ongoing
Classification
Class II
Quantity
2 units
Official record key
device-enforcement:Z-1945-2024

Official wording

Reason: Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances

Code information: UDI: SAT140: (01)05404013814015(11)190507(21)SAT140 SAT136: (01)05404013801138(11)201105(21)SAT136 Serial Numbers: SAT136, SAT140 PTS-11 versions from PTS-11.1 and PTS-12 versions, with C230 accelerator

Distribution pattern: Worldwide - US Nationwide distribution in the state of Virginia and the country of India.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances