Device product

Galaxy System REF GAL-001 (specifically GAL-019, Wired Controller) The Galaxy System and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.

Z-1734-2024

March 09, 2024

Class II

Product summary

Firm
Noah Medical
Event
Event 94193
Status
Ongoing
Classification
Class II
Quantity
17 wired controllers
Official record key
device-enforcement:Z-1734-2024

Official wording

Reason: Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope.

Code information: Model Number: GAL-001 Wired Controller Model Number: GAL-019 UDI-DI code: 00850048825048 (additional information pending) Serial Numbers: GWC-1529 GWC-1538 GWC-1541 GWC-1576 GWC-1579 GWC-1560 GWC-1566 GWC-1586 GWC-1590 GWC-1581 GWC-1589 GWC-1552 GWC-1572 GWC-1547 GWC-1571 GWC-1564 GWC-1537

Distribution pattern: U.S. Nationwide distribution in the states of CA, IN, MD, MO, OH, and PA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope.