Skip to content
Recall Observatory FDA recall evidence

Device product

Prucka 3 Amplifier Model #P1801PA used with CardioLab AltiX electrophysiology recording system and ComboLab AltiX hemodynamic and electrophysiology recording system; and Field Replacement Units spare part #5875569 (Assy CLABIII AMP 128CH 100-240V 50-60HZ).

Z-2258-2024

April 22, 2024

Class II

Product summary

Firm
GE Medical Systems Information Technologies Inc
Event
Event 94717
Status
Ongoing
Classification
Class II
Quantity
145 units
Official record key
device-enforcement:Z-2258-2024

Official wording

Reason: Capacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms.

Code information: UDI N/A for the Prucka 3 Amplifier-not a medical device. CardioLab and ComboLab Systems GTIN #00195278507044 and #00195278507051. Prucka 3 Amplifier serial numbers: SVS23400020SA, SVS23400024SA, SVS23360017SA, SVS23400003SA, SVS23470003SA, SVS23350005SA, SVS23330008SA, SVS23350001SA, SVS23350003SA, SVS23360018SA, SVS23520022SA, SVS23350008SA, SVS23280009SA, SVS23350021SA, SVS23470016SA, SVS23470010SA, SVS23360013SA, SVS23280002SA, SVS23400026SA, SVS23400039SA, SVS23400021SA, SVS23330006SA, SVS23360014SA, SVS23360016SA, SVS23460020SA, SVS23470022SA, SVS23480007SA, SVS23360010SA, SVS23350024SA, SVS24020002SA, SVS24020004SA, SVS24020005SA, SVS24020006SA, SVS24020007SA, SVS24020011SA, SVS23460002SA, SVS23350004SA, SVS23350006SA, SVS23350007SA, SVS23480003SA, SVS23400008SA, SVS23210005SA, SVS23360007SA, SVS23360002SA, SVS23360012SA, SVS23470004SA, SVS23520018SA, SVS23520020SA, SVS24020016SA, SVS23460001SA, SVS23470006SA, SVS23470008SA, SVS23350002SA, SVS23350009SA, SVS23350025SA, SVS23360003SA, SVS23360005SA, SVS23360006SA, SVS23360011SA, SVS23400023SA, SVS23400030SA, SVS23400031SA, SVS23400032SA, SVS23400037SA, SVS23400038SA, SVS23460007SA, SVS23400015SA, SVS23350011SA, SVS23350017SA, SVS23470017SA, SVS23470018SA, SVS23470020SA, SVS23460013SA, SVS23460014SA, SVS23480002SA, SVS23480005SA, SVS23480010SA, SVS23480011SA, SVS23480013SA, SVS23480015SA, SVS23520012SA, SVS23280008SA, SVS23290004SA, SVS23330004SA, SVS23350014SA, SVS23350022SA, SVS23350023SA, SVS23460006SA, SVS23400022SA, SVS23350012SA, SVS23350016SA, SVS23480004SA, SVS23470007SA, SVS23400025SA, SVS23460003SA, SVS23460004SA, SVS23350015SA, SVS23320002SA, SVS23470013SA, SVS23350020SA, SVS23330005SA, SVS23470009SA, SVS23470011SA, SVS23470019SA, SVS23470012SA, SVS23460016SA, SVS23290001SA, SVS23350013SA, SVS23460017SA, SVS23480012SA, SVS23480016SA, SVS23330007SA, SVS23360008SA, SVS23360020SA, SVS23400005SA, SVS23400009SA, SVS23400010SA, SVS23400019SA, SVS23460008SA, SVS23280005SA, SVS23400004SA, SVS23460005SA, SVS23400034SA, SVS23400035SA, SVS23400036SA, SVS23400033SA, SVS23400018SA, SVS23470005SA, SVS23400028SA, SVS23400029SA, SVS23360019SA, SVS23400002SA, SVS23360004SA, SVS23400007SA, SVS23400011SA, SVS23400006SA, SVS23350018SA, SVS23330003SA, SVS23290003SA, SVS23290002SA, SVS23330009SA, SVS23350019SA, SVS23360015SA, SVS23470001SA, and SVS23470002SA.

Distribution pattern: Worldwide - US Nationwide distribution in the states of AK, AZ, CA, CO, FL, GA, IL, KY, MA, ME, MN, MO, NC, NJ, NY, OH, OR, PA, TN, TX, and WA. The countries of Belgium, Canada, Czech Republic, France, Germany, Ireland, Israel, Italy, Japan, Republic of Korea, and Singapore.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Capacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms.