Device product

Product not distributed in the US. Implantable Cardioverter Defibrillators: a. VIVA QUAD XT CRT-D, Model Number DTBA2QQ. b. CLARIA MRI CRT-D SureScan, Model Number DTMA2D1. c. CLARIA MRI QUAD CRT-D SureScan, Model Number DTMA2QQ. d. AMPLIA MRI CRT-D SureScan, Model Number DTMB2D1. e. AMPLIA MRI CRT-D SureScan, Model Number DTMB2D4. f. AMPLIA MRI QUAD CRT-D SureScan, Model Number DTMB2Q1. g. AMPLIA MRI QUAD CRT-D SureScan, Model Number DTMB2QQ. h. COMPIA MRI QUAD CRT-D SureScan, Model Number DTMC2QQ. i. EVERA MRI XT DR SureScan, Model Number DDMB2D4. j. VISIA AF MRI XT VR SureScan, Model Number DVFB2D1. k. VISIA AF MRI S VR SureScan, Model Number DVFC3D4. l. VISIA AF MRI XT VR SureScan, Model Number DVFB2D4.

Z-2212-2024

April 16, 2024

Class II

Product summary

Firm
Medtronic Inc.
Event
Event 94674
Status
Ongoing
Classification
Class II
Quantity
82 units
Official record key
device-enforcement:Z-2212-2024

Official wording

Reason: Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.

Code information: a. GTIN: 00763000612269, Serial Numbers: BLG628370S, BLG628372S, BLG628373S. b. GTIN: 00763000206284, Serial Number: RPO606047S. c. GTIN: 00763000206222, Serial Numbers: RPB629895S, RPB629900S, RPB629902S, RPB629906S, RPB629909S, RPB629912S. d. GTIN: 00763000206307, Serial Number: RPQ610208S. e. GTIN: 00763000206291, Serial Numbers: RPK614721S, RPK614728S. f. GTIN: 00763000206253, Serial Numbers: RPP607499S, RPP607504S, RPP607517S, RPP607518S, RPP607522S, RPP607586S. g. GTIN: 00763000206246, Serial Numbers: RPG643295S, RPG643318S; GTIN 00763000611835, Serial Numbers: RPG643198S, RPG643199S, RPG643200S, RPG643203S, RPG643206S, RPG643209S, RPG643211S, RPG643213S, RPG643218S, RPG643219S, RPG643220S, RPG643221S, RPG643231S, RPG643251S. h. GTIN: 00763000206260, Serial Numbers: RPF648668S, RPF648672S, RPF648674S, RPF648675S, RPF648676S, RPF648677S, RPF648681S, RPF648684S, RPF648685S, RPF648686S, RPF648691S, RPF648693S, RPF648702S, RPF648758S, RPF648760S, RPF648761S, RPF648762S, RPF648763S, RPF648768S, RPF648769S, RPF648770S, RPF648771S, RPF648844S, RPF648846S, RPF648847S. i. GTIN: 00763000206147, Serial Numbers: PGZ663023S, PGZ663203S, PGZ663208S, PGZ663211S, PGZ663216S, PGZ663219S, PGZ663241S, PGZ663264S. j. GTIN: 00763000612191, Serial Numbers: CWH606546S, CWH606548S, CWH606551S, CWH606552S, CWH606553S. k. GTIN: 00763000206352, Serial Numbers: PMX622050S, PMX622052S, PMX622055S, PMX622056S, PMX622057S, PMX622059S. l. GTIN: 00763000206338, Serial Numbers: PLX622054S, PLX622056S, PLX622072S.

Distribution pattern: Worldwide - US Nationwide distribution in the states of CA, CT, FL, IN, MA, MI, NC, NH, NJ, NM, NY, OH, TX, VA and the countries of Australia, Austria, Bulgaria, Canada, China, Croatia, Czech Republic, France, Georgia, Germany, Greece, Guadeloupe, Iceland, Ireland, Italy, Jordan, Kosovo, Latvia, Moldova, Republic Of, Netherlands, Northern Ireland, Norway, Oman, Panama, Poland, Portugal, Qatar, Reunion, Slovakia, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    manufacturing defect