Device product
Hookup Model/Catalog Number: DSD-110-HU0163 Product Description: Hookups provide the interface between DSD Automated Endoscope Reprocessors ( AER ) and endoscopes, which allow perfusion of liquids dispensed by the AER to contact and high-level disinfect the channels of flexible endoscopes. Component: N/A
Z-2539-2024
Product summary
- Event
- Event 94987
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 1082
- Official record key
device-enforcement:Z-2539-2024
Official wording
Reason: IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection
Code information: Lot Code: UDI-DI 00677964064623; Lot Numbers 403142, 403490, 406396, 415623, 420136, 431090, 434467, 439578, 440970, 449929, 453997, 456734, 457984, 460959, 465381, 476333, 480953, 485936, 502488, 505441, 515820, 518697, 530332, 542927, 553320, 554244, 555701, 579924, 585282, 589722, 596830, 607311, 608007, 614507, 618935
Distribution pattern: Worldwide distribution - United States Nationwide and the countries of Canada, Belgium, Czech Republic, Denmark, Germany, Spain, Italy, Slovenia, Slovakia, Australia, Israel, Jordan, Kuwait, Namibia, Saudi Arabia, South Africa, Cote d Ivoire, Afghanistan, Albania, Belarus, Botswana, Kazakhstan, Malaysia, Qatar, Ukraine, United Arab Emirates, Nicaragua, and Nigeria.
Derived failure modes
-
Unknown
IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection