Skip to content
Recall Observatory FDA recall evidence

Device product

TabloCart with Prefiltration Drawer PN-0006813, with power cords: PN-0005746 (10-foot), PN-0006578 (15-foot), PN-0005747 (20-foot)10-, 15- and 20-foot cords respectively), used with the Tablo Hemodialysis System.

Z-2150-2024

April 30, 2024

Class II

Product summary

Firm
Outset Medical, Inc.
Event
Event 94654
Status
Ongoing
Classification
Class II
Quantity
431
Official record key
device-enforcement:Z-2150-2024

Official wording

Reason: A cart with prefiltration drawer, an optional hemodialysis system accessory, has a cord with a female connector that has internal contact pins that may become damaged that could create localized heat, sparks, smoke, melting or charring (thermal event) of the female connector or the associated connector power entry module, which may lead to transient delay in treatment or blistering to the user.

Code information: UDI-DI: 00850001011303. All PN-0006813 serial numbers with power cords: PN-0005746, PN-0006578, PN-0005747

Distribution pattern: US Nationwide distribution in the states of FL, CA, NC, GA, MD, DE, AK, OK, MA, WV, IN, KY, AZ, LA, NE, KS, NJ, CT, NM, CO, SC, TX, PA, NY, OH.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A cart with prefiltration drawer, an optional hemodialysis system accessory, has a cord with a female connector that has internal contact pins that may become damaged that could create localized heat, sparks, smoke, melting or charring (thermal event) of the female connector or the associated connector power entry module, which may lead to transient delay in treatment or blistering to the user.