AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC, 8401 Southern Blvd, Boardman, OH 44512. Usage: Hospital Bed.
Z-0270-2013
Product summary
- Event
- Event 63083
- Status
- Terminated
- Classification
- Class II
- Quantity
- 342 Beds
- Official record key
device-enforcement:Z-0270-2013
Official wording
Reason: The "Retaining washers & Clevis Pins" provided with the AC -powered adjustable hospital beds may fail under certain circumstances. The pins at the top of the leg were disengaging from the channel. This may cause the bed frame to come off tracks or collapse.
Code information: Model #TL500; Serial Number: 001, 002, 003, 004, 005, 006, 007, 010, 013, 014, 015, 016, 017, 018, 021, 022, 023, 024, 025, 026, 027, 028, 029, 030, 031, 032, 033, 034, 035, 039, 040, 041, 042, 043, 044, 045, 046, 047, 048, 049, 050, 052, 053, 054, 055, 056, 057, 058, 059, 060, 061, 062, 063, 064, 065, 070, 075, 081, 09350014, 09350015, 09350016, 09350017, 09350018, 09350021, 09350023, 09350025, 09350026, 09350027, 09350028, 09350029, 09350030, 09350031, 09350032, 09350033, 09350035, 09350038, 09350040, 09350041, 09350042, 09350045, 09350048, 09350049, 09350054, 09350055, 09350056, 09350057, 09350058, 09350059, 09350060, 09350062, 09350063, 09350064, 09350065, 09350066, 09350067, 09350068, 09350069, 09350070, 09350072, 09350074, 09350082, 09350085, 09350087, 09350088, 09350089, 09350090, 09350091, 09350092, 09430001, 09430002, 09430003, 09430004, 09430005, 09430006, 09430007, 09430008, 09430010, 09430013, 09430014, 09430015, 09430017, 09430019, 09430020, 09430021, 09430022, 09430024, 09430026, 09430027, 09430028, 09430029, 09430030, 09430032, 09430033, 09430034, 09430036, 09430037, 09430038, 09430039, 09430040, 09430041, 09430042, 09430043, 09430044, 09430045, 09430046, 09430047, 09430049, 09430050, 09430051, 09430052, 09430053, 09430054, 09430056, 09430057, 09430058, 09430059, 09430060, 09430061, 09430062, 09430063, 09430065, 09430066, 09463067, 09430068, 09430069, 09430070, 09430071, 09430072, 09430074, 09430075, 09430076, 09430077, 09430078, 09430079, 09430080, 09430081, 09430083, 09430085, 09430086, 09430087, 09430088, 09430089, 09430091, 09430093, 09430094, 09430097, 09430099, 09430100, 09520001, 09520002, 09520003, 09520004, 09520005, 09520006, 09520007, 09520008, 09520009, 09520011, 09520012, 09520013, 09520015, 09520017, 09520018, 09520019, 09520020, 09520021, 09520022, 09520025, 09520026, 09520027, 09520028, 09520029, 09520030, 09520031, 09520032, 09520033, 09520034, 09520035, 09520036, 09520037, 09520038, 09520039, 09520040, 09520041, 09520042, 09520043, 09520044, 09520045, 09520046, 09520047, 09520048, 09520049, 09520050, 10040001, 10040003, 10040004, 10040005, 10040006, 10040007, 10040008, 10040009, 10040010, 10040011, 10040012, 10040013, 10040014, 10040015, 10040016, 10040017, 10040028, 10040029, 10040030, 10040032, 10040033, 10040034, 10040035, 10040036, 10040037, 10040038, 10040039, 10040040, 10040041, 10040042, 10040043, 10040044, 10040045, 10040046, 10040047, 10040048, 10040049, 10040050, 10040051, 10040052, 10040053, 10040054, 10040055, 10040056, 10040057, 10040058, 10040059, 10040062, 10040063, 10040064, 10040065, 10040066, 10040067, 10040068, 10040069, 10040070, 10040071, 10040073, 10040074, 10040075, 10040076, 10040077, 10040078, 10400029, 10400050, 10400056, 10400057, 10400058, 10400059, 10400065, 10370016, 10370017, 10370018, 10370019, 10370020, 10370032, 10370033, 10370034, 10370035, 10370038, 10370052, 10370053, 10370054, 10370056, 10370060, 10370063, 10370078, 10370079, 10370080 , 09380031, 09380029, 09380064, 09380070, 09380089, 09380040, 09380038, 09380085, 09380045, 09380054, 09380060, 09380059, 09380058, 09380048, 09380042, 09380038, 09380069, 09380063, 09380067 & 09380065.
Distribution pattern: Nationwide distribution: USA including states of: AZ, IL, KY, MD, MI, OH, OR, PA, TX and VA.
Derived failure modes
-
Unknown
The "Retaining washers & Clevis Pins" provided with the AC -powered adjustable hospital beds may fail under certain circumstances. The pins at the top of the leg were disengaging from the channel. This may cause the bed frame to come off tracks or collapse.