Skip to content
Recall Observatory FDA recall evidence

Device product

AW-M0001 SONARMED MONITOR. Used in conjunction with SonarMed Acoustic Sensors for airway monitoring.

Z-1535-2024

March 25, 2024

Class I

Product summary

Firm
SonarMed Inc
Event
Event 94302
Status
Ongoing
Classification
Class I
Quantity
145 units
Official record key
device-enforcement:Z-1535-2024

Official wording

Reason: Software anomaly resulted in failure to detect a partial obstruction in 2.5 mm sensors and up to 3mm distal to the sensor.

Code information: Model No. AW-M0001; GTIN: 00851334007001; Serial No. AW0572, AW0573, AW0574, AW0577, AW0578, AW0579, AW0580, AW0581, AW0584, AW0585, AW0586, AW0588, AW0591, AW0593, AW0594, AW0600, AW0606, AW0607, AW0615, AW0616, AW0624, AW0626, AW0627, AW0648, AW0650, AW0651, AW0658, AW0664, AW0672, AW0674, AW0675, AW0678, AW0680, AW0681, AW0684, AW0685, AW0701, AW0702, AW0703, AW0704, AW0705, AW0706, AW0707, AW0708, AW0709, AW0710, AW0711, AW0712, AW0718, AW0719, AW0720, AW0721, AW0722, AW0723, AW0724, AW0725, AW0726, AW0727, AW0728, AW0729, AW0730, AW0731, AW0732, AW0733, AW0734, AW0735, AW0736, AW0737, AW0738, AW0739, AW0740, AW0741, AW0742, AW0743, AW0744, AW0745, AW0746, AW0747, AW0748, AW0749, AW0763, AW0764, AW0765, AW0766, AW0767, AW0768, AW0769, AW0770, AW0771, AW0772, AW0773, AW0774, AW0775, AW0776, AW0777, AW0778, AW0779, AW0780, AW0781, AW0782, AW0783, AW0784, AW0785, AW0786, AW0787, AW0788, AW0789, AW0790, AW0791, AW0792, AW0793, AW0794, AW0795, AW0796, AW0797, AW0798, AW0799, AW0800, AW0804, AW0805, AW0814, AW0817, AW0818, AW0819, AW0820, AW0821, AW0825, AW0826, AW0827, AW0828, AW0830, AW0831, AW0832, AW0833, AW0834, AW0835, AW0836, AW0837, AW0838, AW0839, AW0840, AW0841, AW0842, AW0854, AW0855.

Distribution pattern: _AK, AZ, CA, FL, IL, KY, MD, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, & WA_

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    Software anomaly resulted in failure