Skip to content
Recall Observatory FDA recall evidence

Device product

Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967

Z-1321-2024

February 09, 2024

Class II

Product summary

Firm
Preat Corp
Event
Event 94032
Status
Ongoing
Classification
Class II
Quantity
561 units
Official record key
device-enforcement:Z-1321-2024

Official wording

Reason: Due to a manufacturing issue, the screw seat location on the abutment product may be too high, causing mating screw to have a shorter engagement length into the associated implant.

Code information: Model Number: 9000967 UDI-DI code: 00842092165867 Lot Numbers: 235650, 234294, 233667, 233885, 233439, and 231489

Distribution pattern: U.S. Nationwide distribution in the states of AZ, OH and PA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Due to a manufacturing issue, the screw seat location on the abutment product may be too high, causing mating screw to have a shorter engagement length into the associated implant.