Skip to content
Recall Observatory FDA recall evidence

Device product

Abbott Rapid Dx N America LLC Reproductive Endocrinology Control Set McKesson Consult" Pregnancy (hCG) Testing Positive Level / Negative Level 2 X 4 mL

Z-1312-2024

January 09, 2024

Class II

Product summary

Firm
Mckesson Medical-Surgical Inc. Corporate Office
Event
Event 94024
Status
Ongoing
Classification
Class II
Quantity
3 units
Official record key
device-enforcement:Z-1312-2024

Official wording

Reason: Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.

Code information: Catalogue Number: 5011; UDI/DI: 612479202788; Lot No. KN00589;

Distribution pattern: US Nationwide distribution in the states of AZ, PA and TX.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.