Skip to content
Recall Observatory FDA recall evidence

Device product

Sysmex XE-2100C Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #053-2311-4 The Sysmex XE-2100C is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.

Z-0256-2013

June 01, 2012

Class II

Product summary

Firm
Sysmex America, Inc.
Event
Event 62580
Status
Terminated
Classification
Class II
Quantity
69 units
Official record key
device-enforcement:Z-0256-2013

Official wording

Reason: There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.

Code information: catalog ##053-2311-4, all serial numbers

Distribution pattern: Nationwide distribution: USA including Puerto Rico.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.