Skip to content
Recall Observatory FDA recall evidence

Device product

AXIOM MULTIX MT - Intended Radiographic X ray Table is a patient table used in conjunction with the Siemens Multix PRO and TOP Systems radiographic devices Material Number: 8395399

Z-2180-2024

May 31, 2024

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 94777
Status
Ongoing
Classification
Class II
Quantity
50 units
Official record key
device-enforcement:Z-2180-2024

Official wording

Reason: Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire

Code information: UDI-DI: N/A Serial Numbers: 1003 1032 1039 1060 1065 1071 1076 1082 1083 1085 1089 1092 1104 1170 1174 1181 1183 1194 1304 1305 1309 1310 1315 1318 1323 1325 1331 1348 1359 1382 1395 1398 1423 1443 1444 1447 1449 1452 1453 1468 1498 1500 1508 1509 1512 1513 1515 1516 1525 1531

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire