Skip to content
Recall Observatory FDA recall evidence

Device product

Trilogy Evo, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients.

Z-1504-2024

March 19, 2024

Class I

Product summary

Firm
Philips Respironics, Inc.
Event
Event 94263
Status
Ongoing
Classification
Class I
Quantity
76,713 units
Official record key
device-enforcement:Z-1504-2024

Official wording

Reason: Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.

Code information: Model No. AU2110X15B, BL2110X15B, BL2200X15B, BR2110X18B, BR2200X18B, CA2110X12B, CA2200X12B, DE2100X13B, DE2110X13B, DE2200X13B, DS2000X11B, DS2100X11B, DS2110X11B, DS2200X11B, EE2100X15B, EE2110X15B, EE2200X15B, ES2100X15B, ES2110X15B, ES2200X15B, EU2100X15B, EU2100X19, EU2110X15B, EU2110X19, EU2200X15B, FR2100X14B, FR2110X14B, FR2200X14B, FX2100X15B, FX2200X15B, GB2110X15B, GB2200X15B, IA2100X15B, IA2110X15B, IN2100X15B, IN2100X19, IN2110X15B, IN2200X15B, IT2100X21B, IT2110X21B, IT2200X21B, JP2100X16B, JP2110X16B, KR2110X15B, KR2200X15B, LA2100X15B, LA2110X15B, LD2110X23B, ND2100X15B, ND2110X15B, ND2200X15B, RDS2110X11B, SP2100X26B, TR2110X15B, TR2200X15B, UDS2110X11B, VT2110X24B; UDI-DI: 606959051980, 606959055520, 606959055476, 606959052307, 606959054110, 606959051942, 606959055544, 606959054080, 606959054981, 606959055537, 606959055506, 606959054127, 606959055490, 606959051959, 606959055421, 606959051973, 606959055483, 606959055452, 606959055513, 606959060517, 606959055551, 606959060500; All units except those with software version 1.05.06.00.

Distribution pattern: Domestic: Nationwide Distribution; Foreign: AE, AR, AT, AU, BD, BE, BG, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PR, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, US, UY, VN, YT, ZA, ZW.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.