STA Satellite Automated Multi-Parameter Analyzer This system is a fully automatic clinical instrument indicated and intended for the performance of tests on human plasma, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.

Z-0252-2013

June 15, 2012

Class II

Product summary

Firm
Diagnostica Stago, Inc.
Event
Event 63270
Status
Terminated
Classification
Class II
Quantity
232 units US; 5 units Canada
Official record key
device-enforcement:Z-0252-2013

Official wording

Reason: Following a single customer complaint that incorrect INR values were occasionally printed for PT tests, even though the primary units (seconds) were correct, the firm investigated and confirmed the potential for printing incorrect secondary units when using an external printer and the customer printout mode on the STA Satellite¿.

Code information: K082248 D089976 Ref 58104 units

Distribution pattern: Worldwide Distribution - USA (nationwide) and Canada

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Following a single customer complaint that incorrect INR values were occasionally printed for PT tests, even though the primary units (seconds) were correct, the firm investigated and confirmed the potential for printing incorrect secondary units when using an external printer and the customer printout mode on the STA Satellite¿.