Colonoscope, Model Number CF-Q180AL.

Z-1208-2024

January 11, 2024

Class II

Product summary

Firm
Olympus Corporation of the Americas
Event
Event 93832
Status
Ongoing
Classification
Class II
Quantity
2
Official record key
device-enforcement:Z-1208-2024

Official wording

Reason: Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.

Code information: Model Number: CF-Q180AL; UDI DI: 04953170307706; Serial Number: 2602060, 2603763;

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.