Skip to content
Recall Observatory FDA recall evidence

Device product

Sechrist Monoplace Hyperbaric Chamber with Electronic Control Module Model 3600E/ER The Hyperbaric Chamber is intended to provide 100% oxygen to the patient at greater than ambient pressures

Z-0209-2013

August 28, 2008

Class II

Product summary

Firm
Sechrist Industries Inc
Event
Event 52796
Status
Terminated
Classification
Class II
Quantity
65 units total (61 units in the US)
Official record key
device-enforcement:Z-0209-2013

Official wording

Reason: Sechrist conducted a field upgrade on the hyperbaulic chambers equipped with Electronic Control Modules in order to correct a software bug that could potentially affect the performance of the chamber.

Code information: Serial Numbers: 360001-360049, 360051-360064, 360068.

Distribution pattern: Worldwide Distribution - USA (nationwide) and the countries of Bulgaria, and Canada

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software bug