Device product
NautilusTM ECMO Oxygenator with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Product Number: (1) 48145; (2) 48145E
Z-1309-2024
Product summary
- Firm
- Mc3 Inc
- Event
- Event 94016
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 299 units
- Official record key
device-enforcement:Z-1309-2024
Official wording
Reason: May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer
Code information: GTIN: (1) 10854916006888; (2) 10854916006901,20854916006908 (New GTIN (USA only)After reboxing). All Serial Numbers between 481022923 and 481024039
Distribution pattern: US Nationwide distribution.
Derived failure modes
-
Unknown
May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer