Skip to content
Recall Observatory FDA recall evidence

Device product

BD Phoenix NMIC-306 Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene panel (tray) containing control wells and substrate wells with dried biochemical substrates that use traditional qualitative microbiology methods such as fermentation, oxidation, degradation, and hydrolysis in combination with chromogenic and fluorogenic substrates along with carbon sources for bacterial identification and quantitation.

Z-1398-2024

February 29, 2024

Class II

Product summary

Firm
Becton Dickinson & Co.
Event
Event 94180
Status
Ongoing
Classification
Class II
Quantity
11,800 eaches
Official record key
device-enforcement:Z-1398-2024

Official wording

Reason: Products were found to have confirmed drug cross contamination prior to release may lead to incorrect, inadequate, or absent antibiotic treatment or delay in results

Code information: UDI-DI: (10) 3318700 (17)241107 (30)0025 (01) 30382904492923 Lot Number: 3318700 Exp. Date: 07-NOV-2024

Distribution pattern: Worldwide distribution - US Nationwide and the countries of AR, BE, CO, IN, JP, PE, PY, SG, TW, UY.

Derived failure modes

  • Foreign material or chemical contamination

    reason.foreign_material_chemical_contamination · v1.0.0

    contamination prior to release may lead