Device product

MINI STICK MAX 5F X 10 CM STD .018 SS/SS ECHO 1.6" PG- Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-774

Z-1345-2024

January 30, 2024

Class II

Product summary

Firm
Angiodynamics, Inc.
Event
Event 94035
Status
Ongoing
Classification
Class II
Quantity
155 units
Official record key
device-enforcement:Z-1345-2024

Official wording

Reason: Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Code information: UDI: 15051684023177 UPN: H965457741 Lot Number: 5787804 5787805 5800397 5803524 5805070 5805071

Distribution pattern: Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case