Skip to content
Recall Observatory FDA recall evidence

Device product

Centurion Manual surgical kits labeled as: a) STERILE S/B WIRE SCISSOR, Product Code 65140; b) SHARP/SHARP SCISSORS 41/2", Product Code 65840; c) STERILE WIRE CUTTER (WCS144), Product Code 66240; d) STERILE LISTER BANDAGE SCISSOR (LS122), Product Code 66855; e) STERILE UTILITY SHEARS (USH176), Product Code 67225; f) STERILE ANGLED BANDAGE SCISSOR (LIST55), Product Code TRI66670

Z-1449-2024

February 23, 2024

Class II

Product summary

Firm
MEDLINE INDUSTRIES, LP - Northfield
Event
Event 94136
Status
Ongoing
Classification
Class II
Quantity
84224 units
Official record key
device-enforcement:Z-1449-2024

Official wording

Reason: Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.

Code information: a) 65140, UDI/DI 60653160008380 (case) 00653160008388 (each), Lot Numbers: 2023060790, 2023101090; b) 65840, UDI/DI 60653160010208 (case) 00653160010206 (each), Lot Numbers: 2023022790, 2023041190; c) 66240, UDI/DI 60653160008793 (case) 00653160008791 (each), Lot Numbers: 2023112790; d) 66855, UDI/DI 50653160007188 (case) 00653160007183 (each), Lot Numbers: 2023022490, 2023041990, 2023102690; e) 67225, UDI/DI 50653160001766 (case) 00653160001761 (each), Lot Numbers: 2023031690, 2023111490; f) TRI66670, UDI/DI 10653160317975 (case) 20653160317972 (each), Lot Numbers: 2023031490, 2023040690, 2023041090, 2023062790, 2023062890, 2023072490, 2023101690, 2023110190, 2023110990

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Panama, Canada.

Derived failure modes

  • Labeling or packaging

    reason.labeling_packaging · v1.0.0

    weak seal
  • Sterility assurance

    reason.sterility_assurance · v1.0.0

    breach of sterility