Skip to content
Recall Observatory FDA recall evidence

Device product

Colonoscope, Model Number PCF-H190TL.

Z-1212-2024

January 11, 2024

Class II

Product summary

Firm
Olympus Corporation of the Americas
Event
Event 93832
Status
Ongoing
Classification
Class II
Quantity
1
Official record key
device-enforcement:Z-1212-2024

Official wording

Reason: Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.

Code information: Model Number: PCF-H190TL; UDI DI: 04953170420702; Serial Number: 2110150;

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.