Device product
BD BBL Sensi Disc Cefepime - 30 ug¿, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231696¿
Z-1018-2024
Product summary
- Event
- Event 93789
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 2,363,168 total units
- Official record key
device-enforcement:Z-1018-2024
Official wording
Reason: There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
Code information: Catalog No. 231696¿ UDI-DI 30382902316962 Lots 2339439 3118297 3163524 3223410 3243084 3282873 2063393 2119624 2119625 2119626 2147720 2213567 2273732
Distribution pattern: Domestic distribution nationwide. International distribution worldwide.
Derived failure modes
-
Unknown
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.