Synapse PACS - Version 7.2.200

Z-1272-2024

January 10, 2024

Class II

Product summary

Firm
FUJIFILM Healthcare Americas Corporation
Event
Event 94090
Status
Ongoing
Classification
Class II
Quantity
12 Units (12 US)
Official record key
device-enforcement:Z-1272-2024

Official wording

Reason: Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.

Code information: Software Version: 7.2.200, UDI: (01)854904006008(10)0702200

Distribution pattern: Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Australia, Belgium, Brazil, Canada, Colombia, Ecuador, France, Germany, India, Italy, Mexico, Namibia, Netherlands, Portugal, Qatar, Spain, Thailand, Turkey, UAE, United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.