PS LIPPED TIBIAL INSERT ASSY DURACON Intended for knee replacement Product Number: 6742-1-413
Z-0924-2024
Product summary
- Event
- Event 93687
- Status
- Ongoing
- Classification
- Class II
- Quantity
- Unavailable
- Official record key
device-enforcement:Z-0924-2024
Official wording
Reason: Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
Code information: GTIN: 04546540324115 Lot Numbers: 559469, 561926, 584958
Distribution pattern: Nationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand, Norway, and Spain.
Derived failure modes
-
Unknown
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.