Device product
Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LABOR & DELIVERY PACK, kit number HILD52F; b) LABOR & DELIVERY PACK, kit number HILD52G
Z-0829-2024
Product summary
- Event
- Event 93656
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 114 units
- Official record key
device-enforcement:Z-0829-2024
Official wording
Reason: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
Code information: a) HILD52F, UDI/DI none, Lot Numbers: 846221, exp. 6/3/2023; 807221, exp. 7/12/2023; 959231, exp. 11/30/2023; b) HILD52G, UDI/DI none, Lot Numbers: 885231, exp. 4/25/2024
Distribution pattern: US Nationwide distribution.
Derived failure modes
-
Unknown
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.