Device product

EVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002

Z-0775-2024

December 29, 2023

Class I

Product summary

Firm
Datex-Ohmeda, Inc.
Event
Event 93715
Status
Ongoing
Classification
Class I
Quantity
12 units
Official record key
device-enforcement:Z-0775-2024

Official wording

Reason: GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root cause for these elevated levels of formaldehyde. While these test conditions are not representative of typical clinical use conditions,GE HealthCare is taking this action to further reduce the potential for patient exposure.

Code information: UDI/DI ********, Serial Numbers: CBCS01269, CBCT02328, CBCT03050, CBCT02306, CBCT02301, CBCT02303, CBCT02302, CBCT02088, CBCT02309, CBCT02304, CBCU00717, CBCU00559

Distribution pattern: Worldwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root cause for these elevated levels of formaldehyde. While these test conditions are not representative of typical clinical use conditions,GE HealthCare is taking this action to further reduce the potential for patient exposure.