S5 Mast Roller Pump 150, Item Number: 10-88-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.

Z-0274-2013

September 19, 2012

Class II

Product summary

Firm
Sorin Group USA, Inc.
Event
Event 63383
Status
Terminated
Classification
Class II
Quantity
2 units
Official record key
device-enforcement:Z-0274-2013

Official wording

Reason: Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determin

Code information: Serial Numbers: 50E00680--50E00681, 50E00688, 50E00696-50E00706, 50E00708-50E00718.

Distribution pattern: Worldwide distribution: USA (nationwide) including states of: AZ, GA, MI, MS, NC, NY, PA, SC, and TX; and countries of: Austria, Australia, Belgium, Canada, People's Republic of China, Columbia, Denmark, Finland, Germany, India, Iran-Islamic Republic of, Ireland, Japan, Lebanon, Morocco, Pakistan, Poland, Russian Federation, Saudia Arabia, Spain, South Africa, Thailand, Turkey and United Kingdom.

Derived failure modes

  • Potency or specification failure

    reason.potency_specification_failure · v1.0.0

    out of specification