Skip to content
Recall Observatory FDA recall evidence

Device product

VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.

Z-1070-2024

January 09, 2024

Class I

Product summary

Firm
Ventec Life Systems, Inc.
Event
Event 93806
Status
Ongoing
Classification
Class I
Quantity
250 breathing circuits
Official record key
device-enforcement:Z-1070-2024

Official wording

Reason: The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.

Code information: Box Label - Part Numbers: PRT-01013-000 Pouch Label - Part Number: PRT-00802-001 UDI-DI Code: 00855573007594 Lot Number: 220922 Pack Slip Lot Number: 257337

Distribution pattern: U.S.: CA, CO, FL, HI, IN, MO, NC, and WA O.U.S.: Japan

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.