Device product
EVIS EXERA b BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-XT190
Z-0193-2024
Product summary
- Event
- Event 93052
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 719 devices
- Official record key
device-enforcement:Z-0193-2024
Official wording
Reason: There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
Code information: Model BF-XT190 UDI-DI: 4953170402470
Distribution pattern: US Nationwide distribution
Derived failure modes
-
Unknown
There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.