Device product
EZDilate Wire Guided Balloon 8.5-9.5-10.5. indicated for endoscopic dilation of strictures in the alimentary tract in adults and adolescents (>12 years) Model: BD-410X-1055
Z-0940-2024
Product summary
- Event
- Event 93508
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 2991 units
- Official record key
device-enforcement:Z-0940-2024
Official wording
Reason: Product balloon issues: inflating, deflating and/or retrieving the devices; reports of bursting and leaking and events involving foreign body in patient and prolonged procedure.
Code information: UDI-DI:00821925033238 Lot Numbers: 361663, 361664, 365174, 374533, 374534, 375148, 378070, 386123, 386124, 386395
Distribution pattern: Nationwide. Foreign: AU BR BR CA DE HK IN JP SG
Derived failure modes
-
Foreign material or chemical contamination
foreign body