Device product

STAT-Site M BHB Test Strips, used with STAT-SITE M Beta-HB Meter as an in vitro diagnostic for blood ketone values.

Z-0490-2024

October 02, 2023

Class II

Product summary

Firm
Stanbio Laboratory, LP
Event
Event 93300
Status
Ongoing
Classification
Class II
Quantity
2,351
Official record key
device-enforcement:Z-0490-2024

Official wording

Reason: Deterioration in the stability of the Test Strips results in diminished low-range detection capability and inability to meet linearity performance claims. The root cause of the stability deterioration in is identified as the change in material that caused a pH shift in the coating solution resulting in precipitation during coating runs. The device defect can cause false negative results that may lead to a delay in treatment of patients with associated conditions such as diabetic ketoacidosis.

Code information: UDI-DI: 00657498000090, Lot/Expiration: 203044/2024-01-31, 209644/2024-06-30

Distribution pattern: Worldwide - US Nationwide distribution including in the states of AZ, RI, MO, UT, AK, LA, TX, OH, GA, MN, NY, CA, NJ, IL, WA, KS, MI, NC, PA, MA, CO, KY, NM, FL, MD, VA, IN, AL, SD, MS, WV, WI, ID, MT, WY and the countries of Canada and Singapore.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Deterioration in the stability of the Test Strips results in diminished low-range detection capability and inability to meet linearity performance claims. The root cause of the stability deterioration in is identified as the change in material that caused a pH shift in the coating solution resulting in precipitation during coating runs. The device defect can cause false negative results that may lead to a delay in treatment of patients with associated conditions such as diabetic ketoacidosis.