Skip to content
Recall Observatory FDA recall evidence

Device product

Patient Information Center iX with Software Version Number 4.x in the following: (1) Patient Information Center iX Product Number: 866389; (2) Patient Information Center iX Expand Product Number: 866390; (3) PIC iX Essentials Product Number: 867093; (4) PIC iX Essentials Expand Product Number: 867154

Z-2608-2023

August 17, 2023

Class II

Product summary

Firm
Philips North America Llc
Event
Event 92992
Status
Ongoing
Classification
Class II
Quantity
960 units
Official record key
device-enforcement:Z-2608-2023

Official wording

Reason: Patient Information Center (PIC iX) Release 4.x Surveillance Crash Caused by Intel Graphics Driver Error-with a blank screen and subsequently require a manual reboot of the hardware to restart and continue central patient monitoring. Potential for a delay in the detection of a change or deterioration in the condition of one or more patients

Code information: UDI: (01)00884838104594(10)4.0.1 (01)00884838104594(10)4.0.2 (01)00884838112070(10)4.1.0 Software Version Number 4.x

Distribution pattern: Nationwide Foreign: Austria Bahrain Belgium Canada Czech Republic Denmark Finland France Germany Greece India Ireland Italy Japan Jordan Kenya Kuwait Lebanon Lithuania Netherlands Norway Poland Portugal Qatar Romania Slovakia South Africa South Korea Spain Sweden Sweden Switzerland United Kingdom Utd.Arab.Emir. Uzbekistan

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Patient Information Center (PIC iX) Release 4.x Surveillance Crash Caused by Intel Graphics Driver Error-with a blank screen and subsequently require a manual reboot of the hardware to restart and continue central patient monitoring. Potential for a delay in the detection of a change or deterioration in the condition of one or more patients